Dacryocystorhinostomy (DCR) Laser Multidiode OFT 4G
October 8, 2022Automated Visual Field Analyzer M700
The Medmont M700 leverages over 30 years of commercial experience to bring practitioners and patients an effective, economical and reliable tool for assessing visual fields.
Full Field Coverage (160°)
Advanced Fast Threshold Testing Strategy, Employing Baysean Testing Techniques
Flicker Test Facility, With Proven Early Field Loss Detection Capabilities
Patient Reliability Indicators
- False Positives
- False Negatives
- Fixation Losses
- Video Fixation Eye Tracking Monitor
- Automatic Fixation Tracking
Field Analysis Tools
- Glaucoma Progression
- Pattern Defect Index (PD)
- Overall Defect Index (OD)
- Cluster Analysis (Glaucoma) Index
- Regression Analysis Over Entire Field or Localized Areas
- Baseline Analysis
Display Options
- Grey Scale
- Numeric Decibel Data
- Patient Hill of Vision Deviation (Localized Defect Identification
- Age Normal Deviation (General Depression Identification)
- Severity of Loss Indicators (Relative Probability of Loss)
Result Outputs
- Customizable Page Layout
- Single Result Per Page
- Left Eye/Right Eye On One Page
- Difference Maps
- Multiprint with 5 Results Per Page (Extendable)
Regression Graphs Printing
- Pattern Defect Index
- Overall Defect Index
- Data Histograms
- OP
- 3D Display
Size
- Stimulator Unit Dimensions:
- 626mm Wide x 438mm Deep x 713mm High
- Stimulator Unit Weight:
- 12kg
- Shipping Dimensions / Weight:
- 71cm x 52cm x 85cm, 23 kg (Unit and Box)
Performance
- Stimulator Screen:
- Part Hemispherical Bowl Radius 30cm Integrated Diffusing Surface
- Test Fields:
- Binocular 30°/40° : 21-128 Points Binocular Driving 80° : 119 Points Central 22A 22° : 45 to 96 Points Central 30° : 100 Points Driving Test 50°/80° : 103 Points Flicker 15°/22° : 69 Points Full 50° : 164 Points Glaucoma 22°/50° : 104 Points Macula 10° : 49 Points Neurological 50° : 164 Points Peripheral 30° to 50° : 73 Points Flash Scan 22°/30° : 40 Points Spatially Adaptive 50° : 39 to 168 Points CV% 100 Points 60° : 100 Points Central 22 22° : 50 Points
- Stimulus Source:
- Rear Projection Light Emitting Diode
- Stimulus Color:
- Pale Green 565nm Half Bandwidth 28nm Red 626nm (Optional)
- Stimulus Size:
- Goldmann Size III (0.43°), Model CR Red (0.72°) (Optional)
- Stimulus Intensity:
- 16x3dB Levels 0db (Max Brightness) to 45dB (Min Brightness) +/-1dB
- Stimulus Duration:
- Adjustable: 0.1 to 9.9 sec. (nom. 0.2 sec)
- Patient Response Time:
- Adaptive to Patient Speed, Operator Selection of Normal or Slow Ranges Adjustable: 0.1 to 9.9 sec (nom. 1.1 sec)
- Minimum Inter-stimulus Delay:
- Adjustable 0.1 to 9.9 sec (nom. 0.4 sec)
- Background Illumination:
- 10 asb (3.2cd/m2), Automatic Level Control 31.5 asb (10cd/m2) – German Driving Test
- Test Lens Diameter:
- 38mm
- Fixation Method:
- Automatic Iris Tracking During Test with Relaxed Tolerance. Heijl-Krakau Blind Spot Method. Visual and Audible Warning of Fixation Errors. Video Fixation Monitoring
Electrical
- Stimulator Unit Power:
- 100-240 VAC 50-60Hz , 0.25-0.15A
Computer Minimum Requirements
- PC and Mains Powered Peripherals:
- EN/IEC60950 Compliant
- PC Requirements OS:
- Microsoft™ Windows™ 10 Professional (recommended) and Home (x64 and x86), or, Windows Server 2016, 2019 (64 bit editions only)
- Processor: Intel™ Generation 6 i5 or later
- Motherboard: Genuine Intel™ chipset highly recommended
- VIA chipsets have proven to be unreliable
- Memory: 8 GB for non-video captures, 16 GB for video captures.
- Hard Disk Space: 40 GB for non-video, 200GB for video captures (the more the better)
- Video cards: GPUs with dedicated memory of 2 GB or more are recommended. GPUs that share memory require 128 MB of memory minimum and are not recommended.
- Screen resolution: 1920 x 1080 recommended, minimum supported 1280 x 720
- Printer: Bubble-jet or Laser-jet printer (Color / Monochrome)
- USB
- At least 1 USB 2.0 Compliant port on the PC